Wattmeter (Dynamometer) & Power Measurement

Test Instrument

Measures real power by multiplying voltage and current inside one instrument: a dynamometer movement whose torque tracks V×I×cosφ. Home of the two-wattmeter method and half the numericals in the Measurements paper.

What it looks like

Wattmeter (Dynamometer) & Power Measurement

A typical test instrument — exact shape, colour and markings vary between manufacturers.

Types & variants

Dynamometer (electrodynamic) analogLow power factor (LPF) wattmeter (transformer tests)Unity PF (UPF) wattmeterDigital wattmeter / power analyzerEnergy meter (kWh — the integrating cousin, see Heavy)Plug-in power meters (kill-a-watt style)

Key specs

Current coil rating

A

The fixed coil in series with the load: 5/10/20 A ranges. Overload burns it — always start on the highest range.

Pressure coil rating

V

The moving coil across the load: 150/300/600 V ranges via multipliers.

Power factor class

UPF / LPF

LPF wattmeters give full deflection at cosφ = 0.2 — mandatory for transformer open-circuit tests where PF is tiny and a UPF meter barely moves.

Multiplier constant

×

Reading × range factor = watts; getting this wrong is the classic lab error.

Connection error

Pressure coil on supply side vs load side each mis-measures slightly (coil consumption) — a standard viva question with a compensation-winding answer.

Markings

Terminals marked M L (current coil: Mains–Load) and C V (pressure coil: Common–Voltage), with ± polarity dots; ranges printed by each terminal pair. Reversed deflection in the two-wattmeter method means that meter's reading is negative.

Standard values

Lab standards: 5/10 A, 150/300/600 V dynamometer wattmeters, UPF and LPF versions, class 0.5. Digital: single- and three-phase power analyzers reading W, VA, VAR, PF, THD.

How to choose

1) Routine loads: UPF dynamometer or digital meter. 2) Transformer OC test: LPF type, or the reading is unreadably small. 3) Three-phase: two-wattmeter method needs two matched meters (total P = W₁+W₂, and tanφ = √3(W₁−W₂)/(W₁+W₂) — the formula to memorise). 4) Modern work: a power analyzer adds PF and harmonics for SMPS/VFD loads that fool analog meters.

Pinout & package

Four (or five) terminals: current coil in series (M→L), pressure coil across (C→V). Portable wooden/bakelite lab cases with range selectors; digital units use the same wiring or clamp CTs.

Example circuits

  • Single-phase load: wattmeter + voltmeter + ammeter, deriving PF = W/(V×I)
  • Two-wattmeter method on a 3-phase induction motor, watching W₂ go negative below PF 0.5
  • Transformer OC test with LPF wattmeter reading iron loss
  • Blondel's theorem in practice: n−1 wattmeters for n wires

Common failures

Burnt current coil from overload (meter reads zero forever), bent pointer, insulation breakdown between coils, and in labs mostly wiring errors: swapped M/L, missing ± link, wrong multiplier factor applied.

How to test

Known resistive load check: a 500 W heater on 230 V should read ~500 W and match V×I. Verify each range's multiplier with the same load. Compare against a digital power meter for calibration drift.

Substitutes

Digital power analyzers replace dynamometers everywhere except the syllabus itself. For rough audits, a clamp power meter or plug-in energy monitor. The energy meter (kWh) integrates what the wattmeter reads instantaneously — related, not interchangeable.